User’s Manual
BIM1000 Series
External Control
Control and Monitor
Starting and monitoring measurements
When starting measurements with internal triggers
When a measurement is completed, the next measurement starts automatically. The following figure
shows a timing chart including the INDEX (pin 22) and EOM (pin 21) signal outputs when the trigger
source
is set to INTERNAL.
Item
Time
T2: Delay
The time set with delay
. 5 ms minimum.
T3: Measurement time
Measurement time × average count
Measurement time: Time set with the sampling speed
.
Average count: Average count set with the average function
.
T4: Computation time
0.3 ms
T5: EOM signal output (HOLD)
When EOM
is set to HOLD. Holds low level until the next measure-
ment.
T6: EOM signal output (PULSE)
When EOM
is set to PULSE. Outputs a pulse for the specified time.
T7: Judgment delay
0.2 ms
Measurement
ON
OFF
For HOLD
ON
OFF
T2
T3
T4
T5
Delay
Computation
Judgment
result output
For PULSE
OFF
T6
ON
Delay
T7
INDEX
EOM
Internal trigger
Measurement start