Operation Verification
Verification Tests
Operation Verification
The procedures in this section are used to provide a high degree of confidence
that the Switch Module is meeting published specifications. The Operation
Verification tests are usually a subset of the Performance Verification tests and
are suitable for checkout after performing repairs.
For the Keysight E1463A Switch Module, Operation Verification is performed by
completing the Closed Channel Resistance Test as described in the Performance
Verification test procedures (Test 2-1). This test is usually sufficient to verify that
the instrument is meeting its specifications.