Driver Installation and Commands
3
V3500A User’s Guide
55
Basic V3500A commands
Command Name
Reset
Description
This command resets the V3500A to its factory
power- on state.
Syntax
*RST<LF>
Response
OK
Command Name
Triggered power read
Description
This is the basic measurement command for the V3500A.
Syntax
*TRG<LF>
Response
<Value>
Return value
Returns the present V3500A reading.
Comments
The present read cycle of the V3500A is interrupted, and a
new one is started. The measurement from the new cycle
is returned.
Command Name
Non-triggered power read
Description
This command returns the present reading of the V3500A.
Syntax
PWR? <LF>
Response
<Value>
Comments
This is a non- triggered measurement. If the input power has
changed during the present measurement cycle, the measured
power may be unpredictable.
Command Name
Zero command
Description
This command zeroes the V3500A.
Syntax
ZERO<LF>
Response
OK
Comments
This internal correction routine removes offset errors in the
measurement. This is necessary to preserve accuracy when
measuring low- level signals.
Содержание V3500A
Страница 1: ...Keysight V3500A Handheld RF Power Meter User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X V3500A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...XIV V3500A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...XVI V3500A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...XVIII V3500A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 49: ...Driver Installation and Commands 3 V3500A User s Guide 29 ...
Страница 52: ...32 V3500A User s Guide 3 Driver Installation and Commands ...
Страница 53: ...Driver Installation and Commands 3 V3500A User s Guide 33 ...
Страница 103: ...RF Measurement Basics 4 V3500A User s Guide 83 V0 Vpeak 2 ...