Operations and Features
3
Keysight U8001A/U8002A User’s and Service Guide
41
Programming Over Current Protection
Over current protection guards the load against output currents reaching values
greater than the programmed protection level.
The following steps show how to enable and disable the over current protection
(OCP), how to set the OCP trip level and how to clear the over voltage condition.
To set the OCP trip level and enable OCP
1 Press on the “Over Current” button.
– The LIMIT annunciator will be turned off.
– The OCP annunciator will fast-blink continuously.
– The display will show the OCP value and the character “A” will fast-blink
continuously.
2 Turn the knob to ad just the OCP value.
3 Press on the “Over Current” button again to end the adjustment and enable
the OCP.
– Display will show “donE”.
– The OCP annunciator will be turned on.
NOTE
To cancel the operation, allow the unit to idle for five seconds.
Содержание U8001A
Страница 8: ...8 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 69: ...Service Guide 5 Keysight U8001A U8002A User s and Service Guide 69 Figure 5 4 Graph of load transient response ...