Keysight U8001A/U8002A User’s and Service Guide
Constant Current Operation
The following steps show you how to perform the constant current operation.
1 Short the binding posts.
– Press on the “Power” button to turn off the power supply.
– Connect a short circuit between the positive (+) and negative (—) binding
posts.
2 Turn on the power supply.
– Press on the “Power” button to turn on the power supply. The power supply
will then perform a self-test.
– The display turns on all segments and shows the firmware version briefly
after that.
– The output is disabled by default.
– The OFF annunciator turns on.
3 Set the display to the limit mode.
– Press on the “Display Limit” button in order to turn on the LIMIT
annunciator.
– The display will show the voltage and current limit values.
4 Ad just for the desired voltage limit.
– Press on the “Voltage/Current” button. When seeing the voltage value, “V”
blinks, turn the knob to adjust for the desired voltage limit value.
NOTE
When you press on the Display Limit button, the voltage and current limit values
will be shown on the display for about five seconds. If there is no activity
detected, the display will return to meter view mode.
Содержание U8001A
Страница 8: ...8 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 69: ...Service Guide 5 Keysight U8001A U8002A User s and Service Guide 69 Figure 5 4 Graph of load transient response ...