![Keysight U4421A Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/keysight/u4421a/u4421a_hardware-and-probing-manual_1961143028.webp)
4
Setting up an E5381A Differential Flying Lead Probe Configuration for U4421A Module
28
Keysight U4421A MIPI D-PHY Analyzer and Exerciser Hardware and Probing Guide
The following table maps the channel number of the flying leads with their color coding and lists the
D-PHY signal connection to be probed by each of these cables.
D-PHY Signals to Flying Leads Mapping
In a E5381A probe, there are:
• Two flying leads for probing D-PHY clock positive and negative signals.
• Two flying leads for probing D-PHY data lane 0 positive and negative signals.
• One flying lead each for probing D-PHY data lane 1, 2, and 3, if used.
Channel Number of
the flying lead
Color of the flying lead
D-PHY Signal Connection
Positive (light grey) side of the
flying lead
Negative (dark grey) side of the flying
lead
0
Black
D0p
GND
1
Brown
D0p
D0n
2
Red
D0n
GND
3
Orange
D1p
GND
4
Yellow
D1p
D1n
5
Green
D1n
GND
6
Blue
Not used
Note used
7
Purple
Not used
Not used
8
Black
D2p
D2n
9
Brown
D2n
GND
10
Red
D3p
GND
11
Orange
D3p
D3n
12
Yellow
D3n
GND
13
Green
CLKp
GND
14
Blue
CLKp
CLKn
15
Purple
CLKn
GND
Clk
White
D2p
GND