Environmental Conditions
This instrument is designed for indoor use and in an area with low condensation.
The table below shows the general environmental requirements for this
instrument.
Environmental condition
Requirement
Temperature
Operating condition
– 0 °C to 55 °C
Storage condition
– –40 °C to 70 °C
Humidity
Operating condition
– Relative humidity up to 80% RH at 40 °C (non-condensing)
– Proportional to 50% RH from 40 °C to 55 °C (non-condensing)
Storage condition
– Up to 95% RH at 40 °C (non-condensing)
Altitude
Up to 2000 m
Pollution degree
2
NOTE
The U3606B complies with the following safety and EMC compliances:
Safety compliance
– IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition)
– Canada: CAN/CSA-C22.2 No. 61010-1-04
– USA: ANSI/UL 61010-1:2004
EMC compliance
– IEC 61326-1:2005/EN61326-1:2006
– CISPR11:2003/EN55011:2007, Group 1 Class A
– Canada: ICES/NMB-001:Issue 4, June 2006
– Australia/New Zealand: AS/NZS CISPR 11:2004