Digital Multimeter Operation
2
Setting the Resolution
U3606A User’s and Service Guide
55
Setting the Resolution
You can select either 4½ digit or 5½ digit resolution for the AC voltage,
DC voltage, AC+DC voltage, AC current, DC current, AC+DC current,
2- wire resistance, and 4- wire low- resistance measurements.
•
5½ digit readings have the best accuracy and noise rejection.
•
4½ digit readings provide for faster readings.
•
The range and resolution of the frequency, pulse width, and duty cycle
measurements follows the configuration of the AC voltage or the AC
current function.
•
Continuity and diode tests have a fixed 4½ digit display.
•
Capacitance measurements have a fixed 3½ digit display.
Key
Description
Press
Shift
>
4½ 5½
to toggle between 4½ digit and 5½ digit mode.