2
Digital Multimeter Operation
Making Measurements
26
U3606A User’s and Service Guide
Measuring DC voltage
1
Press
to make DC voltage measurements. (The DC annunciator is
shown on the display.)
2
Connect the red and black test leads to input terminals
V
(red) and
LO
(black) respectively as shown in
.
3
Probe the test points and read the display.
Table 2-1
DC voltage measurement summary
Item
Description
Available ranges
19.9999 mV, 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 1000.00 V
Measurement method
Sigma Delta A-to-D converter
Input impedance
10 M
Ω
± 2% range (typical) in parallel with capacitance < 120 pF
Input protection
1000 V
rms
on all ranges, < 0.3 A short circuit
V
DC annunciator
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...