Characteristics and Specifications
8
DC Power Supply Specifications
U3606A User’s and Service Guide
257
Supplementary characteristics
Front panel resolution
Voltage
1 mV
Current
0.1 mA
Transient response time
Less than 450 ms for output to recover to within 50 mV following a
change in output current from full load to half load or vice versa.
Command processing time
Average time for output voltage to begin to change after receipt of
digital data when instrument is connected directly to the USB or GPIB
is less than 100 ms.
Overvoltage protection (for CC mode):
•
Accuracy: 0.5% + 0.5 V
•
Activation time
[2]
: < 2ms
Overcurrent protection (for CV mode):
•
Accuracy: 0.5% + 0.05 A
•
Activation time
: < 2 ms
[1] Contacts and leads resistance may contribute an additional error of 6 mV/A (typical).
[2] Average time for the detection of OVP or OCP condition.
Table 8-8
DC power supply performance specifications (continued)
Parameter
Specifications
Maximum output
programming range:
Range S1
•
CV/CC: 0 V to 31.500 V/0 A to 1.05 A
•
OV/OC: 31.500 V/1.05 A
•
OVP/OCP: 33.000 V/1.1 A
Range S2
•
CV/CC: 0 V to 8.4 V/0 A to 3.15 A
•
OV/OC: 8.4 V/3.15 A
•
OVP/OCP: 8.8 V/3.3 A
Temperature coefficient:
± (% of offset)
maximum change in
output/readback per °C for
0 ºC to 18 ºC/28ºC to 55 ºC
•
Voltage: 0.005% + 0.5 mV
•
Current: 0.02% +1 mA
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...