6
Calibration Procedures
Calibration Count
188
U3606A User’s and Service Guide
Calibration Count
The calibration count feature provides an independent “serialization” of
your calibrations. You can query the instrument to determine how many
calibrations have been performed. Note that your instrument was
calibrated before it left the factory. When you receive your instrument, be
sure to read the count to determine its initial value.
The calibration count increments up to a maximum of 32767 after which
it rolls over to 0. Since the value increments by one for each calibration
point, a complete calibration may increase the value by many counts.
The calibration count is stored in the nonvolatile memory, and does not
change when power has been turned off, after a Factory Reset (
*RST
command), or after an Instrument Preset (
SYSTem:PRESet
command).
Reading the calibration count
1
Unsecure the instrument for calibration. See
for more information on how to unsecure
the instrument.
2
Press
MinMax
to view the calibration count. The calibration count is
displayed in the upper secondary display.
3
The instrument will return to the calibration operation after a few
seconds of inactivity.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...