Verification and Performance Tests
Recommended Test Equipment
144
U3606A User’s and Service Guide
Connecting the current monitoring resistor
To eliminate the output measurement errors caused by voltage drops in
the leads and connections, connect the current monitoring resistor
between the
output terminal and the load as a four terminal device.
Connect the current monitoring leads inside the load- lead connections
directly at the monitoring points on the resistor element.