Keysight U3401A User’s and Service Guide
DC Current Verification Test
1
Connect the calibrator to the front panel Hi and Lo input connectors.
2
Select each function and range in the order shown in
. Provide the
input shown in
3
Make a measurement and observe the result. Compare measurement results
to the appropriate test limits shown in
. (Be certain to allow for
appropriate source settling when using the Fluke 5520A).
Table 4-2
DC voltage verification test
Function
Input
Range
Error from nominal one year
DC voltage
0.00 V
500 mV
±40
μ
V
0.0000 V
5 V
±0.4 mV
0.000 V
50 V
±4 mV
0.00 V
500 V
±40 mV
0.0 V
1000 V
±400 mV
450.00 mV
500 mV
±130
μ
V
4.5000 V
5 V
±1.3 mV
45.000 V
50 V
±13 mV
450.00 V
500 V
±130 mV
900.0 V
1000 V
±580 mV
CAUTION
Set the calibrator output to 0 V before disconnecting it from the multimeter
input terminals.