1
Calibration Procedures
36
Keysight U1270 Series Service Guide
Unsecuring the Instrument for Calibration
Before you can adjust the instrument, you must unsecure it by entering the
correct security code.
The default security code is set to 1234.
To unsecure the instrument from the front panel
1
Power-on the instrument and press the
and
keys simultaneously
for more than 1 second.
The calibration security code entry is shown on the display.
Figure 1-5
Unsecuring the instrument
2
Key in the default security code if you are unsecuring your instrument for the
first time.
– Press
or
to move the cursor to the right or to the left.
– Press
or
to increment or decrement the digit.
3
Press
when you are done.
NOTE
If you forget your security code, see
“To reset the calibration security code to its
Shift
View
Esc
Auto
Trig
Hold
Peak
MaxMin
Auto
Range
Exit
Dual
Setup
Hz % ms
Log
Содержание U1271A
Страница 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Страница 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 53: ...Service and Maintenance 2 Keysight U1270 Series Service Guide 53 Figure 2 1 Testing Fuse 1 Smart Smart ...
Страница 54: ...2 Service and Maintenance 54 Keysight U1270 Series Service Guide Figure 2 2 Testing Fuse 2 Smart Smart ...
Страница 56: ...2 Service and Maintenance 56 Keysight U1270 Series Service Guide Figure 2 3 Replacing Fuse 1 1 2 3 ...
Страница 57: ...Service and Maintenance 2 Keysight U1270 Series Service Guide 57 Figure 2 4 Replacing Fuse 2 1 2 3 ...