5
Performance Tests and Calibration
64
Keysight U1241AN User’s and Service Guide
Input Connections
Test connections to the multimeter are best accomplished using the K-type
thermocouple wire and mini-connectors for temperature measurement. The
J-type thermocouple wire and mini-connectors can also be used for temperature
measurements. Shielded, twisted-pair, Teflon interconnect cables of minimum
length are recommended between the calibrator and the multimeter. Cable
shields should be earth ground referenced. This configuration is recommended for
optimal noises and settling time performance during calibration.
Performance Verification Tests
The performance verification tests are recommended as acceptance tests when
you first receive the multimeter. The acceptance test results should be compared
against the one year test limits. After acceptance, you should repeat the
performance verification tests at every calibration interval.
If the multimeter fails performance verification, adjustment or repair is required.
Adjustment is recommended at every calibration interval. If no adjustment is
being performed, it is highly recommended to establish a ‘guard band’, using not
more than 80% of the specifications, as the verification limits.
NOTE
Users are highly recommended to read the
before performing performance verification tests.
Содержание U1241AN
Страница 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Страница 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 28: ...1 Getting Started 28 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...