:TRIGger Commands
29
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
957
:TRIGger:PATTern
Command Syntax
:TRIGger:PATTern <pattern>
<pattern> ::= <string>[,<edge_source>,<edge>]
<string> ::= "nn...n" where n ::= {0 | 1 | X | R | F} when
<base> = ASCii
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $} when
<base> = HEX
<edge_source> ::= {CHANnel<n> | NONE}
<n> ::= 1 to (# of analog channels) in NR1 format
<edge> ::= {POSitive | NEGative}
The :TRIGger:PATTern command specifies the channel values to be used in the
pattern trigger.
In the <string> parameter, each bit corresponds to a channel as described in the
following table:
The format of the <string> parameter depends on the :TRIGger:PATTern:FORMat
command setting:
• When the format is ASCii, the string looks just like the string you see on the
oscilloscope's front panel, made up of 0, 1, X (don't care), R (rising edge), and F
(falling edge) characters.
• When the format is HEX, the string begins with "0x" and contains hex digit
characters or X (don't care for all four bits in the nibble).
With the hex format string, you can use the <edge_source> and <edge>
parameters to specify an edge on one of the channels.
You can only specify an edge on one channel. When an edge is specified, the
:TRIGger:PATTern:QUALifier does not apply.
Query Syntax
:TRIGger:PATTern?
Oscilloscope Models
Value and Mask Bit Assignments
2 analog channels only
Bits 0 and 1 - analog channels 2 and 1.
NOTE
The optional <edge_source> and <edge> parameters should be sent together or not at all. The
edge can be specified in the ASCII <string> parameter. If the edge source and edge
parameters are used, they take precedence.
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...