![Keysight P9241 Скачать руководство пользователя страница 910](http://html1.mh-extra.com/html/keysight/p9241/p9241_scpi-programmers-manual_1961682910.webp)
910
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
29
:TRIGger Commands
:TRIGger:HOLDoff:RANDom
Command Syntax
:TRIGger:HOLDoff:RANDom {{0 | OFF} | {1 | ON}}
The :TRIGger:HOLDoff:RANDom command enables or disables the random trigger
holdoff mode. This mode randomizes the holdoff time from one acquisition to the
next. The randomized holdoff time values will be between the values specified by
the :TRIGger:HOLDoff:MINimum and :TRIGger:HOLDoff:MAXimum commands.
The random trigger holdoff mode ensures that the oscilloscope re-arms after each
acquisition in a manner that minimizes or eliminates the likelihood of triggering at
the beginning of a DDR burst. Randomizing the holdoff time increases the
likelihood that the oscilloscope will trigger on different data phases of a
multi-phase (8 data transfer) burst. This mode mixes up the traffic pattern the
oscilloscope triggers on and is very effective when used on repeating patterns.
Query Syntax
:TRIGger:HOLDoff:RANDom?
The :TRIGger:HOLDoff:RANDom? query returns random trigger holdoff mode
setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also
•
•
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...