92 Keysight N9923-90001 User’s Guide
DTF (Distance to Fault) Measurements
DTF Measurement Settings
—
More
then
DTF (Lin)
Faults are displayed on the Y-axis in linear
(unitless) format.
DTF Start and Stop Distance
In DTF measurements, you set the physical length of cable or other device to
be tested. The FieldFox calculates the frequency range of the measurement
from this distance. The longer the cable to be tested, the lower the frequencies
that are used. You can also set the frequencies manually using the Frequency
Mode [Bandpass] setting.
How to set Start and Stop Distance
— With a DTF measurement present, press
Freq/Dist
.
— Then choose from the following:
—
Start Distance
Enter a value using the numeric keypad, the
arrows, or the rotary knob, then
Enter
. By default, the Start Distance
is set to 0 Meters. This means that the measurement will display
faults starting at the point at which calibration standards are
connected.
—
Stop Distance
Enter a value between the start distance and 5 km (or
16,404 ft.) using the numeric keypad, the
arrows, or the rotary
knob, then
Enter
.
Frequency Mode
All DTF measurements are made with frequency settings and, using Inverse
Fourier Transform (IFT), the time and distance to faults are calculated.
How to set Frequency Mode
With a DTF measurement present,
— Press
Meas Setup 4
— Then
Frequency Mode
— Choose one of the following:
—
Lowpass Mode
The frequency range of a DTF measurement is set
automatically based on the Start and Stop Distances. Use Lowpass
mode when the DUT is a cable ONLY.
The start and stop frequencies for the measurement are always annotated
on the screen below the start and stop distances.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...