70 Keysight N9923-90001 User’s Guide
Time Domain – Option 010
Time Domain (Transform) Settings
The DUT is a 100 meter transmission cable. The Line Loss value is.1 dB/meter.
This means that a signal traveling ONE WAY through the cable will lose 10 dB
of power (100 m *.1dB/m). Because the FieldFox performs this measurement
with 1 port, the test signal travels down the cable and then back, for a total
loss of 20 dB.
For the purpose of illustrating this point, connect an OPEN to the end of the
cable – a maximum-sized fault - for 100% reflection of the 300 MHz test
signal.
Without compensation
for the loss of the cable, a –20 dB response would be
visible at 100 meters, which is the OPEN at the end of the DUT. This is from 10
dB of loss through the cable in each direction.
With compensation
for the loss using the manufacturer’s specification, the
FieldFox compensates the trace as though the signal traveling through 100
meters was increased by +20 dB. Therefore the response will show 0 dB for
100% reflection.
Data Chain
This setting, available only on the Table Settings, reverts to
Standard
when the
FieldFox is Preset. Choose from the following:
—
Standard
– The normal FieldFox data processing chain. Transform
calculations are performed AFTER error correction and trace math.
—
8510
– The data processing chain used by the Keysight 8510 network
analyzer. Transform calculations are performed BEFORE error correction
and trace math. Learn more at the FieldFox Supplemental Online Help
website:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...