Keysight N9923-90001 User’s Guide
51
NA (Network Analyzer) Mode
NA Mode Settings
— Then
Average <n>
where <n> is the number of measurements to average.
— Enter a value using the numeric keypad. Enter 1 for NO averaging.
— Press
Enter
.
— Then
Average Mode
Choose from the following:
—
Sweep
- Each data point is based on the average of the same data
point being measured over <n> consecutive sweeps. The average
counter shows the number of previous sweeps that have been
averaged together to form the current trace. When the counter
reaches the specified count, then a ‘running average’ of the last <n>
sweeps is displayed.
—
Point
- Each data point is measured <n> times and averaged before
going to the next data point. On subsequent sweeps, averaging
restarts by measuring each data point again <n> times. The average
counter is not updated because data is not displayed until all the
averages have been applied.
—
Point averaging is usually faster than sweep averaging.
However, you
may need to increase the Point Average count to obtain the same
level of noise reduction as with sweep averaging.
— While averaging is in process, press
Sweep 3
then
Restart
to restart the
averaging at 1.
IF Bandwidth
The FieldFox converts the received signal from its source to a lower
intermediate frequency (IF). The bandwidth of the IF bandpass filter is
adjustable. Reducing the IF receiver bandwidth reduces the effect of random
noise on a measurement. However, narrower IF bandwidths cause longer
sweep times.
How to set IF BW
—
Press BW 2.
— Then
IF BW
— Then choose from the following:
—
10 Hz
|
100 Hz
|
1 kHz
|
10 kHz
|
100 kHz
More
30 Hz
|
300 Hz
|
3 kHz
|
30 kHz
Smoothing
Trace smoothing averages a number of
adjacent
data points to smooth the
peak-to-peak noise values on a displayed trace. The number of adjacent data
points that are averaged is known as the smoothing aperture. Aperture is set
by specifying a percentage of the X-axis span.
Trace smoothing does NOT significantly increase measurement time.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...