40 Keysight N9923-90001 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
S21 and S12 transmission measurements are used to measure the loss or gain
through a DUT over a specified frequency range. Both ends of the DUT must be
connected to the FieldFox. The FieldFox signal source is transmitted out one of
the test port connectors, through the DUT, and into the other test port
connector
How to measure S-parameters
1.
Press
Preset
then
Preset
2.
Press
Measure 1
then choose from the following:
—
S11
Reflection measurement at port 1.
—
S21
Forward 2-port transmission measurement.
—
S12
Reverse 2-port transmission measurement. May require an
option.
—
S22
Reflection measurement at port 2. May require an option.
OR
select a multi-trace configuration. Learn more on
.
3.
Press
Freq/Dist
then either
Start
and
Stop
or
Center
and
Span
to enter a
Frequency Range for the measurement.
4.
Press
BW 2
then
IFBW
to select the IF Bandwidth for the measurement.
Narrower bandwidths require more time to sweep, but lowers trace noise.
5.
Press
Sweep 3
then
Resolution
to select the number of data points for the
measurement. More data points require more time to sweep.
6.
Press
Cal 5
to calibrate the measurement. Learn more in
.
7.
All other settings can be made AFTER calibration without compromising
measurement accuracy.
Mixed-Mode S-Parameters
In NA Mode ONLY, mixed-mode S-parameters (also known as Balanced
measurements) are available with Option 212.
Because the FieldFox has only two test ports, only reflection measurements are
available. Connect the balanced input or output of your DUT to the FieldFox
ports 1 and 2
For highest accuracy, a Full 2-port calibration is required.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...