C: Instrument Calibration
Over time, the active components in the FieldFox age and the performance
may degrade or drift. To ensure that the FieldFox is performing to the
published specifications, you should have an instrument calibration performed
periodically.
How Often Should I Calibrate my Instrument?
You determine the calibration period which best meets your requirements.
However, a 12 month calibration cycle is appropriate for most users.
There are two things to consider: performance drift and connector wear.
The instrument specifications are set to consider the performance drift that
may occur over a 12 month period. Therefore, getting the instrument
calibrated at 12 month intervals ensures that the analyzer maintains
performance within the operating specifications. If you need the analyzer to
maintain more consistent operation, you may want to have the instrument
calibrated more often.
Connector wear is a bigger factor and depends on the number of connections
that are made. The test ports become noticeably worn after 500 to 700
connections. This could represent about 12 months with average use. With
more frequent connections, the calibration cycle should be sooner. You can
extend the time between calibrations and thereby save money by using
connector savers and by performing proper Connector Care. Learn more in
“Connector Care Review” on page 261
How do I Get an Instrument Calibration?
To get your Keysight instrument calibrated, send it to one of the Keysight
Technologies service centers. Please visit this website to learn how:
www.keysight.com/find/calibration
.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...