Keysight N9923-90001 User’s Guide 165
VVM (Vector Voltmeter) Mode - Option 308
VVM Mode Settings
This setting survives a Preset and Mode Preset.
How to set Resolution
— Press
Scale/Amptd
— Then
Resolution
— Then choose:
—
1
0.0 (default)
—
2
0.00
IF Bandwidth
Reducing the IF receiver bandwidth reduces the effect of random noise on a
measurement. However, narrower IF bandwidths cause longer measurement
times.
For highest accuracy, Zero AFTER changing the IF Bandwidth. Learn more in
How to set IF BW
— Press
BW 2
— Then
IF BW
— Then choose from the following:
300 Hz
| 1 kHz
| 3 kHz
| 10 kHz
| 30 kHz
Output Power
Set the power level out of the FieldFox to High, Low, or manually set power
level to a value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss
devices to place the signal farther from the noise floor. However, for devices
that are sensitive to high power levels such as amplifiers, use the Low power
setting.
For best measurement accuracy, use the Manual power setting at -15 dBm.
After calibration, the power level can be decreased for amplifiers, or increased
for higher dynamic range.
How to set Output Power
— Press
Meas Setup 4
.
Power Level settings in this mode will NOT change Power Level settings in
other modes. To help prevent damage to your DUT, use caution when
changing modes with your DUT connected to the FieldFox test ports.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...