Keysight N9923-90001 User’s Guide 149
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
—
Frequency
Sets the center frequency of the carrier on which pulse
modulation resides. This is used to set amplitude correction of the
USB Power Sensor.
How to set Time
Specify the time over which data will be acquired by the USB Power Sensor.
When Trace Graph is selected,
Tr 1
is annotated with these values. Time can
be specified using either of the following pairs of values:
—
Center 1
and
Time/div 1
Specify the center time and time per division.
—
Start Time
and
Time Length
Specify the start time and length of
acquisition.
— A positive start time indicates a delay after the sensor detects a pulse
signal.
— A negative start time displays data that is acquired BEFORE the sensor
detects a pulse signal (or T-zero).
— Follow each by entering a value using the numeric keypad, the
arrows,
or the rotary knob.
— After using the keypad, select a multiplier key. Learn about multiplier
abbreviations in
“Multiplier Abbreviations” on page 34
.
Zoom Window
When Trace Graph is selected as the measurement, a secondary Zoom Trace
can be used to view a zoomed portion of
Tr 1
. When Zoom Window is ON, blue
vertical posts appear on
Tr 1
to indicate the current start, center, and stop
times of the zoom window. The zoom window can be scaled using the standard
Scale settings.
How to make Zoom Window settings
— Press
Measure
— Then
Trace Graph
— Press
Trace 6
— Then
Zoom Window OFF ON
— Then choose from the following:
—
ON
Opens the zoom window on the bottom half of the FieldFox
screen. The window includes the zoomed trace.
—
OFF
Closes the zoom window.
—
Zoom Center
While monitoring the T within the blue vertical posts on
Tr 1
,
enter a value for the center time using the numeric keypad, the
arrows, or the rotary knob.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...