Keysight N9923-90001 User’s Guide 147
Pulse Measurements Mode – Option 330
Supported Power Sensors
Supported Power Sensors
Pulse measurements are made using any of the Keysight U2020 X-Series USB
Peak Power Sensors.
Zeroing of the U2020 X-Series USB Peak Power Sensors is performed
automatically at power up, every 5 seconds, and prior to measuring low level
signals.
Switching between Pulse Measurement Mode and Power Meter Mode
NOT Supported
- Using a single USB Peak Power Sensor and switching
between Power Meter Mode and Pulse Measurement Mode. Both modes
will attempt to program the sensor, which could cause one or both modes
to hang, or at least report inaccurate data.
Supported
- Using one U2000A Average Power Sensor connected to the
top USB port for Power Meter Mode, and one Peak Power Sensor
connected to the bottom USB port for Pulse Measurement Mode.
How to Connect the Power Sensor
Connect your Power Sensor to one of the FieldFox USB ports.
A sequence of status lights are displayed on the base of the power sensor.
—
Green blinking
light indicates the power sensor is communicating or
self-tests are being performed. Because the FieldFox is usually
communicating with the power sensor when it is connected, this is the
normal mode of operation.
—
Amber light
indicates Zeroing is being performed.
—
Red light
indicates an error has occurred with the power sensor. Refer to
power sensor documentation.
—
No light
indicates the FieldFox is not communicating with the power
sensor. This is the case when in Hold mode or when NOT in Pulsed
Measurement Mode.
Use an attenuator between the DUT and the power sensor when
measuring power levels that are higher than +20 dBm. The attenuator
value can be subtracted from the measurement using the Offset feature,
discussed in
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...