Keysight N9923-90001 User’s Guide 143
Frequency Offset Using Power Sensor (FOPS) – Option 208
FOPS Settings
—
Dwell /Point
Specifies a settling time (in seconds) after the internal source
steps to the next frequency and before the power sensor makes a
measurement.
Amplitude Markers
Amplitude markers are horizontal dotted lines that you can move to any
amplitude level on the screen. The two amplitude markers can be set
independently or used as a pair of delta markers. Amplitude markers can be
used as a visual comparison aid with the data trace.
How to make Amplitude Marker settings
— Press
Marker
— Then
More
— Then
Ampl Markers
— Then choose from the following:
—
OFF
Amplitude markers are NOT displayed.
—
ON
Amplitude markers ARE displayed at their previous positions.
— Then
Ampl Mkr 1 or 2
— Enter a value in dBm for the marker using the numeric keypad, the
arrows, or the rotary knob.
— Press
Delta Ampl Mkr
— Then choose from the following:
—
ON
AmplMkr 2 value is relative to AmplMkr 1. When AmplMkr 1 is
moved, AmplMkr 2 also moves to maintain the AmplMkr 2 value
between them.
—
OFF
The two amplitude markers move independent of one another.
Trace Math
Trace Math allows you to compare data traces and perform simple math. Learn
how to use Trace Math in
“All about Trace Math” on page 189
. All Data and
Memory features work as documented EXCEPT for the following:
To Compare Src Power (Input) and Rx Power (Output):
1.
Meas Rx Power
2.
Store to memory
3.
View Data & Memory
4.
Meas Src Power.
When Trace, then Data -> Mem is pressed, if a Src Power trace is active,
then RxPower is actually stored into memory.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...