140 Keysight N9923-90001 User’s Guide
Frequency Offset Using Power Sensor (FOPS) – Option 208
FOPS Settings
—
Gain
Not available until source data is stored into memory. Displays the
Output power / Source power (memory trace).
—
Rx Power
Displays the raw received power at the USB power sensor.
Sweep Type and Frequency
Set the source and receiver frequencies for the FOPS measurements. The
receiver is a Keysight U-Series Power Sensor. These are broadband
measurement devices, which means that they measure power over a very wide
frequency range. Tables of frequency and power correction factors are stored
within the power sensors. The receiver (Rx) frequencies are used ONLY to apply
the appropriate correction factors to the power levels measured with the power
sensor.
How to set Sweep Type and Frequency
— Press
Freq/Dist
— Then
Sweep Type
— Then choose from the following:
—
CW
Used to make standard (non-offset) power meter measurements at a
single frequency. Learn more in
“USB Power Meter Settings” on page 128
—
Swept
Used to make swept FOPS measurements.
——
Then enter source (Src) frequencies as either Center and Span
or Start and Stop. Both methods set the FieldFox internal
source sweep range out the Port 1 RF Output connector.
——The frequency limits are determined by the limits of both the
FieldFox and the power sensor.
——The LOW frequency is limited by the higher of either the
FieldFox or the power sensor minimum frequencies.
——The HIGH frequency is limited by the lower of either the
FieldFox or the power sensor maximum frequencies.
——Then
More
.
——Then
Freq Display
. This determines how the receiver
frequencies below the graticules are displayed. Choose from
the following:
It is YOUR responsibility to ensure that ONLY the signal of interest is
present at the power sensor. All other signals MUST be filtered
appropriately.
To make FOPS measurements at a single frequency, choose Swept, then
enter a Src Center frequency, then Span = 0 Hz.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...