122 Keysight N9923-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
Verifying Calibration and Jumper Cable Integrity
Verifying Calibration and Jumper Cable Integrity
After calibrating, it is important to verify that the calibration is good. When
using a jumper cable, also verify that the cable is of high quality.
Verify a Calibration
— Connect a LOAD standard at the calibration reference plane (where
calibration standards were connected). This may be at the end of the
jumper cable or at FieldFox
PORT 1
connector.
— In CAT Mode, select
Return Loss
.
— In NA Mode, select
S11
with Log Mag format.
— For frequencies
lower than 9 GHz
, observe the trace on the FieldFox
screen:
—Lower than 35 dB indicates a GOOD calibration.
—Higher than 30 dB indicates that you should recalibrate.
Verifying Phase Accuracy
— Connect a LOAD standard at the calibration reference plane (where
calibration standards were connected).
— In NA Mode, select a S11 Reflection with Polar or Smith Chart format.
— Because all LOAD standards have delay, you should see a small amount of
phase rotation as a function of frequency. In general, the measurement
result should agree with the characteristics of the calibration standard.
Test the Jumper Cable
With the LOAD standard still connected, move the jumper cable while
observing the trace.
— If the measurement trace is relatively stable, the jumper cable is of good
quality.
— If you observe significant movement in the peaks of the measurement trace
when moving the cable (>5 dB), the jumper cable may need to be replaced.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...