Keysight N9923-90001 User’s Guide
113
Calibration for NA, CAT, and VVM Modes
Normalize
uses a THRU standard or cable between port 1 and port 2 to cal an
S21 and S12 Transmission measurement (NA Mode) and a 2-port Insertion
Loss measurement (CAT Mode). In VVM Mode, this is performed using Zero.
How to perform a Response Cal
— Select the measurements to be calibrated. See the relevant Mode (NA, CAT,
or VVM) for measurement selections.
— Press
Cal 5
then
Response Cal
— For 1-port measurements:
—Select either
Open Response
or
Short Response
for the port to be
calibrated. The availability of this calibration on Port 2 may require an
option.
—Connect an OPEN or SHORT standard to the specified port and press
Measure
— For 2-port measurements:
—Select
Normalization
—Connect a short, high-quality, phase stable cable between the
FieldFox port 1 and port 2 connectors, then press
Measure
— Press
Finish
.
— Connect the DUT.
CAL ON U
is shown on the screen when a User Cal (Ex: Response Cal) is
correcting ONLY the appropriate measurement. For example, when an Open
Response Cal on Port 2 is performed,
CAL ON U
is shown for an S22
measurement only.
With a Normalization Cal, all subsequent insertion loss measurements are
made relative to the insertion loss of the cable used as the THRU standard.
For example, if you use a cable with 1 dB of loss, then after Normalization,
the display will show 0 dB of loss with this cable in place. Therefore, for
highest accuracy, when measuring the DUT also attach the cable that was
used in the normalization cal.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...