110 Keysight N9923-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
—
Press
Change Gender
to change the gender of the DUT
connector.
—
Then press
Next…
to confirm the selection and continue the
process.
—
For each port:
—
Use the
arrows or rotary knob to select the Cal Kit that
you have, and are going to use, for the specified ports.
—
Press
Next…
to confirm the selection and continue the
process.
—
Then press
Finish
to end the selection process.
4. Calibration Type
- The FieldFox always RECOMMENDS a Cal Type that will
quickly and accurately calibrate all of the displayed S-parameters.
Change from the RECOMMENDED Cal Type ONLY if you understand the
implications. Learn more about Cal Types in
To select a different Cal Type:
—
Press
Change Cal Type
.
—
Then using the
arrows or rotary knob, select a Cal Type,
—
Then press
Select and Finish
.
Begin Calibration
— Press
Start Calibration
.
— Follow the Cal Wizard prompts. Connect the specified standard at the point
where the DUT will be connected, then press
Measure
.
— At any time, press
Back
to re-measure a standard if you feel it was not
properly connected.
— Press
Finish
to complete the calibration.
CAL ON U
is shown on the screen for all displayed measurements that are
corrected with the Mechanical Cal.
ECal
ECal is a complete solid-state calibration solution. Every ECal module contains
electronic standards that are automatically switched into position during a
calibration. These electronic standards have been measured at the factory and
If an error appears (“Failure to compute calibration steps…”), check to
ensure that the frequency range of the Cal Kit covers the frequency range
of the measurement. You can verify the frequency range of your Cal Kit at:
www.keysight.com/find/fieldfoxsupport
. Click Cal Kits.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...