Keysight N9927-90001 User’s Guide 419
Noise Figure Mode (Option 356 – CPU2 Only)
How to Define a Noise Figure Measurement
Frequency Range and Frequency Context
Set the range of frequencies over which you would like to make DUT type
measurements and frequency context for noise measurements.
When the frequency range is changed after a calibration is performed, the cal
becomes interpolated. Learn more in
.
How to set Frequency Range and Frequency Context
— Press
Freq/Dist
(Measurement) or
for Frequency Context:
— Press
Meas Setup
>
DUT)
>
RF (DUT In)
and
IF (DUT Out)
— Then choose from the following
—
Start
and
Stop
frequencies - beginning and end of the sweep.
—
Center
and
Span
frequencies – the center frequency and span of
frequencies (half on either side of center).
— Follow each setting by entering a value using the numeric keypad,
arrows, or the rotary knob.
— After using the keypad, select a multiplier key. Learn about multiplier
abbreviations in
— After use the
arrows or the rotary knob, press
Enter
.
Figure 23-15
Содержание N9913A
Страница 6: ...6 ...
Страница 56: ...56 Keysight N9927 90001 User s Guide CAT Cable and Antenna Test Mode 2 Port Insertion Loss Measurements ...
Страница 82: ...82 Keysight N9927 90001 User s Guide TDR Time Domain Reflectometry Option 215 TDR Measurement Settings ...
Страница 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 142: ...142 Keysight N9927 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Страница 262: ...262 Keysight N9927 90001 User s Guide IQ Analyzer IQA Option 351 CPU2 Only Other Features Available ...
Страница 286: ...286 Keysight N9927 90001 User s Guide Interference Analyzer SA Mode Option 236 Record Playback SA Option ...
Страница 292: ...292 Keysight N9927 90001 User s Guide Reflection Mode SA Models Option 320 Reflection Mode Settings ...
Страница 316: ...316 Keysight N9927 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 332: ...332 Keysight N9927 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 346: ...346 Keysight N9927 90001 User s Guide VVM Vector Voltmeter Mode Option 308 A B and B A Measurements ...
Страница 370: ...370 Keysight N9927 90001 User s Guide AM FM Metrics SA Only and Option 355 Interpreting AM FM Metrics Error Messages ...
Страница 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Страница 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Страница 528: ...528 Keysight N9927 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 544: ...544 Keysight N9927 90001 User s Guide Safety Considerations Certification and Compliance Statements ...
Страница 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Страница 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Страница 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Страница 555: ...555 Keysight Handheld Analyzers N99xxA User s Guide E Keysight Software End User Licensing Agreement EULA ...
Страница 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Страница 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 565: ......