Keysight Technologies N9040B UXA Signal Analyzer Service Guide 75
Instrument Messages
Introduction
Event Message Format
The event messages are listed in numerical order according to their message
number. Advisory messages do not have numbers, and are listed in
alphabetical order.
An explanation is included with each error to further clarify its meaning. Some
errors are specified in industry standards and there are references to the IEEE
Standard 488.2-1992,
IEEE Standard Codes, Formats, Protocols and Common
Commands for Use with ANSI / IEEE Std 488.1-1987
. New York, NY, 1992.
Advisories
Advisory messages tell the front panel user some useful information. (For example, “File saved
successfully” or “Measuring the fundamental”.)
Advisory messages appear in the Status Panel at the bottom of the display. The message remains
until you press a key, or another message is displayed in its place.
Advisory messages are not logged in the error queues.
Grayout messages are a special type of Advisory, which appear when you attempt to access a
function that is not available. This could be a grayed out front panel key, or an inappropriate SCPI
command. There are two types of grayout messages: Benign and Forced.
1. Benign
: the requested function is not available because it does not make sense
with the current instrument settings. Changing it does not affect the current
measurement. (For example, setting the number of FFTs/Span when you are not
in the FFT mode.)
A benign grayout gives an Advisory type of message only when the front panel
key is pressed.
The requested function cannot be changed from the front panel, but it can be
changed remotely.
2. Forced
: the requested function is not available either because changing it
would cause an invalid measurement, or because of hardware limitations, or
because the selection conflicts with other settings. (For example, selecting the
electrical attenuator when the frequency span includes frequencies above 3.6
GHz.)
A forced grayout function cannot be changed either from the front panel or
remotely. It generates a special type of Advisory message. It also only appears
on the front panel when the key is pressed. Remotely, the message will appear
in the event queue as a warning “-221, Settings conflict; <conflict
description>”.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...