Keysight NFA Series Noise Figure Analyzers Service Guide
115
Instrument Messages
Condition Messages
Condition Errors 101 to 199, Measurement Integrity
This series of errors corresponds to the bits in the
STATus:QUEStionable:INTegrity
register. The second column in the table
shows the corresponding bit in that register.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be viewed in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 141 indicates an Input Overload condition has been
detected, error 1129 indicates that failure has been cleared.
Two bits in this register are “summary bits” for registers at a lower level. There
are no error messages associated with these bits; they exist only as status bits,
which can be read with a
STATus:QUEStionable:INTegrity?
event query
or a
STATus:QUEStionable:INTegrity:CONDition?
query.
Note that these summary bits summarize the state and history of the event
registers at the lower level. This is true even for bits in the
STATus:QUEStionable:INTegrity
condition register. This means that:
— The summary bits read by the
STATus:QUEStionable:INTegrity:CONDition?
query are true if any
event bits are set in any of the :INTegrity sub-registers :SIGNal or
:UNCalibrated.
— The summary bits read by the
STATus:QUEStionable:INTegrity?
event
query are true if any event bit has undergone a false-to-true transition with
the PTRansition filter set, or a true-to-false transition with the NTRansition
filter set, in any of the :INTegrity sub-registers :SIGNal or :UNCalibrated
Thus, the summary bits
cannot
be used to determine the current state of a
lower level condition bit; only the state and history of the lower level event bits.
This register is itself summarized as bit 9 of the STATus:QUEStionable register,
as described in the section
“Condition Errors 601 to 699, Error Summaries” on
.
Err#
Bit in
status
register
Message
Error or
Warning
More Information
133
0
Signal Summary
status bit
only
This bit is the summary bit for the
STATus:QUEStionable:INTegrity:SIGNal
sub-register.
135
1
No Result
E
135
1
No Result;Turn on MCE
E
To calculate Timing and Phase results in the
Code Domain Power view of Mod Accuracy, the
"Multi Channel Estimator" must be set to ON.
Otherwise these results are invalid.
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...