26
Chapter 2
Test Descriptions
Input VSWR
Test Setup
Figure 2-3
Input VSWR: Equipment Connection
Test Procedure
NOTE
Throughout the Input VSWR Test Procedure the term ‘the Test Record’
refers to the Input VSWR Test Record detailed in the relevant Appendix
for the model number being tested.
The Input VSWR test requires the following stages:
1. Calibrating the Vector Network Analyzer 1 from 10 MHz to 1.5 GHz
or 3 GHz
2. Measuring VSWR from 10 MHz to 500 MHz
3. Measuring VSWR from 500 MHz to 1000 MHz
4. Measuring VSWR from 1000 MHz to 1500 MHz
5. Measuring VSWR from 1500 MHz to 3000 MHz
6. Calibrating the Vector Network Analyzer 2 from 3 GHz to 6.7 GHz or
26.5 GHz
7. Measuring VSWR from 3000 MHz to 6700 MHz
VECTOR NETWORK
ANALYZER 1
Port 1
Cable 1
Adapter 1
(if Applicable)
Adapter 2
(if Applicable)
Содержание N8972A
Страница 9: ...ix Declaration of Conformity N8972A ...
Страница 10: ...x Declaration of Conformity N8973A ...
Страница 11: ...xi Declaration of Conformity N8974A ...
Страница 12: ...xii Declaration of Conformity N8975A ...
Страница 19: ...Contents xix ...
Страница 20: ...xx Contents ...
Страница 31: ...11 2 Test Descriptions This chapter descibes each test required for calibration and performance verification ...
Страница 108: ...88 Chapter2 Test Descriptions Measurement Jitter ...
Страница 124: ...104 Chapter3 Technical Specifications General Specifications ...
Страница 136: ...116 AppendixA Model N8972A Test Records Measurement Jitter Test Record ...
Страница 148: ...128 AppendixB Model N8973A Test Records Measurement Jitter Test Record ...
Страница 160: ...140 AppendixC Model N8974A Test Records Measurement Jitter Test Record ...
Страница 174: ...154 AppendixD Model N8975A Test Records Measurement Jitter Test Record ...