36
Keysight N774xC Series Multiport Power Meter Programming Guide
3
Instrument Setup and Status
Parameters:
none
Response:
The bit value for the operation enable mask as a
16-bit unsigned integer
value (0 .. +65535)
Example:
:stat:oper:enab?
->
+128
Command:
:STATus[n]:OPERation[:EVENt]?
Syntax:
:STATus[n]:OPERation[:EVENt]?
Description:
Returns the Operational Slot Status Event Register (OSSER) of slot
n
.
Parameters:
none
Response:
The results for the individual slot events (a
16-bit unsigned integer
value, where 0
≤
value
≤
65535):
Bit
4-15
3
1-2
Mnemonic
Not used
Channel
n
: Zeroing ongoing
Not used
Decimal Value
8
Example:
:stat0:oper?
->
+0
Command:
:STATus[n]:OPERation:CONDition?
Syntax:
:STATus[n]:OPERation:CONDition?
Description:
Returns the Operational Slot Status Condition Register of slot
n
.
Parameters:
none
Response:
The results for the individual slot events (a
16-bit unsigned integer
value, where 0
≤
value
≤
65535):
Bit
4-15
3
1-2
Mnemonic
Not used
Channel
n
: Zeroing ongoing
Not used
Decimal Value
8
Example:
:stat0:oper:cond?
->
+0