Installation Note N5247-90106 41
Installation Procedure for the Upgrade
Semirigid Cables Required for Upgrading to an Option 419 PNA
To see images showing the location of these cables, click the Chapter 6
bookmark “Bottom RF Cables, 4-Port, Option 419” in the PDF Service Guide
1
.
New parts are listed in
— W69 (reuse) (N5247-20112) A27 mixer brick to rear-panel EXT TSET DRIVE
LO OUT (J5)
— W68 (N5247-20088) rear panel port RF2 OUT (J12) to W67
* Install SMA load previously removed from A10 26.5 GHz source board
onto rear panel port RF2 OUT (J12).
— W123 (reuse) (N5247-20020) A49 port 2 receiver attenuator to port 2 RCVR
B IN
— W115 (reuse) (N5247-20027) A45 port 2 bias tee to port 2 CPLR THRU
— W46 (N5247-20019) A36 port 2 coupler to port 2 CPLR ARM
— W114 (reuse) (N5247-20034) A41 port 2 source attenuator to port 2
SOURCE OUT
— W112 (N5247-20029) A44 port 4 bias tee to A35 port 4 coupler
— W124 (reuse) (N5247-20046) A49 port 2 receiver attenuator to A27 mixer
brick (B)
— W146 (reuse) (N5247-20058) A32 port 2 reference coupler to REF 2
SOURCE OUT
— W145 (N5247-20066) A27 mixer brick (R2) to REF 2 RCVR R2 IN
, install clamp part number 1400-1334 to secure
W145.
1. See