50 Installation Note N5242-90125
Description of the Upgrade
Installation Procedure for the Upgrade
-
Step 33. Verify the PNA Analyzer Program is Running with the Correct
Options
1.
Start the Network Analyzer program.
2.
Once the Network Analyzer program is running:
—
Press
Help
>
About NA
and verify that Option 205 is listed in the
PNA application.
3.
After successful installation of all upgrades, some features require some
adjustments to ensure the instrument meets its specified performance.
Refer to the following Web site:
http://na.support.keysight.com/pna
Step 34. Perform Post-Upgrade Adjustments and Calibration
Adjustments
The following adjustments must be made due to the hardware changes of the
analyzer.
— source adjustment
— IF gain adjustment
— receiver characterization
— receiver adjustment
— LFE receiver adjustment
— IF Response adjustment (Options S93090xA/B, S93093A/B, or S93094A/B
Only.)
These adjustments are described in the PNA Service Guide and in the PNA
on-line HELP. A list of equipment required to perform these adjustments is also
found in the service guide.
To view this service guide information, click the Chapter 3 bookmark "Tests
and Adjustments" in the PDF Service Guide
1
.
After the specified adjustments have been performed, the analyzer should
operate and phase lock over its entire frequency range.
If the option is still not enabled or if the older option has not been
removed, contact Keysight Technologies. Refer to
1. See