3-34 Keysight N5224B/25B Service Guide
Tests and Adjustments
Performance Tests
3-
— If the analyzer still fails this test, troubleshoot the source section of the
analyzer and then repeat this test. Refer to
“Checking the Source Group” on
.
Frequency Accuracy Test
Function of the Test: To verify the frequency accuracy and range of the
analyzer’s source output.
Specification Tested: Test Port Output–CW Accuracy
Equipment Used:
— Frequency counter
— Test cable
— Adapters
Description of the Test:
This test is performed over the full frequency range of the source synthesizer
board, not the full frequency range of the analyzer. To generate the higher
frequencies, the analyzer passes the synthesizer signal through a series of
frequency doublers. These doublers exactly double the source frequency, so
the deviation from a perfectly accurate frequency is exactly doubled. The
frequency accuracy is specified as the ratio parts per million (ppm), so this ratio
is unaffected by the signal doubling. Therefore, only the frequency accuracy of
the synthesizer board needs to be tested.
1.
The analyzer is Preset.
2.
The analyzer is set up for a CW measurement on Port 1.
3.
A test cable is connected between Port 1 and a frequency counter with
any necessary adapters.
4.
The signal frequency is measured and compared with the analyzer source
frequency setting. The difference must be less than the source frequency
divided by 16 for a 1 part per million (ppm) specification.
5.
This test is repeated at several frequencies across the range of the source
synthesizer board.
If the Analyzer Fails this Test:
— Verify the accuracy of the 10 MHz OCXO by using a frequency counter to
measure the rear-panel 10 MHz REF OUT. If the 10 MHz reference is off by
more than 10 Hz, perform the
“Touchscreen Adjustment and Verification”
and then repeat this test.
Trace Noise Test
Function of the Test: To measure the stability of a signal in the internal source
and receiver system of your analyzer.
Содержание N5224B
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...12 Keysight N5224B 25B Service Guide Contents...
Страница 84: ...3 52 Keysight N5224B 25B Service Guide Tests and Adjustments Adjustments 3...
Страница 145: ...Keysight N5224B 25B Service Guide 4 61 Troubleshooting This page intentionally left blank...
Страница 212: ...5 50 Keysight N5224B 25B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 17 4 Port Receiver Group...
Страница 213: ...Keysight N5224B 25B Service Guide 5 51 Theory of Operation Receiver Group Operation Figure 5 18 2 Port Receiver Group...
Страница 220: ...5 58 Keysight N5224B 25B Service Guide Theory of Operation Power Supply Group Operation 5...
Страница 225: ...Keysight N5224B 25B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure...
Страница 490: ...6 270 Keysight N5224B 25B Service Guide Replaceable Parts Replaceable Parts Listings 6...
Страница 608: ...8 102 Keysight N5224B 25B Service Guide General Purpose Maintenance Procedures Correction Constants 8...
Страница 612: ...Appendix 4 Service Guide EEPROM Address Assignments and Location N5224 5A B PNA and N5244 5A B PNA X Instruments...
Страница 613: ...Installation Note Xxxxx xxxxx 3...