4-52 Keysight N5224B/25B Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
Checking the Receiver Group
Equipment Used for These Tests
For all of the following checks, refer to the block diagrams at the end of this
chapter and to any of the following that are appropriate:
—
“2-Port Configuration, Option 200, S/N Prefix <6021” on page 6-31
—
“Bottom RF Cables, Standard 2-Port Configuration, Option 200, S/N Prefix
—
“2-Port Configuration, Option 205, S/N Prefix <6021” on page 6-46
—
“Bottom RF Cables, 2-Port, Option 205, S/N Prefix <6021” on page 6-49
—
“2-Port Configuration, Option 219, S/N Prefix <6021” on page 6-68
—
“Bottom RF Cables, 2-Port, Option 219, S/N Prefix <6021” on page 6-71
—
“4-Port Configuration, Option 400, S/N Prefix <6021” on page 6-137
—
“Bottom RF Cables, Standard 4-Port Configuration, Option 400, S/N Prefix
—
“Bottom Assemblies, 4-Port, Option 405, S/N Prefix <6021” on page 6-153
—
“Bottom RF Cables, 4-Port, Option 405, S/N Prefix <6021” on page 6-156
—
“4-Port Configuration, Option 419, S/N Prefix <6021” on page 6-180
—
“Bottom RF Cables, 4-Port, Option 419 (Ports 1 and 2), S/N Prefix <6021”
—
“Bottom RF Cables, 4-Port, Option 419 (Ports 3 and 4), S/N Prefix <6021”
—
“Bottom Assemblies, 4-Port, Option 420, S/N Prefix <6021” on page 6-192
—
“Bottom RF Cables, 4-Port, Option 420 (Ports 1 and 2), S/N Prefix <6021”
—
“Bottom RF Cables, 4-Port, Option 420 (Ports 3 and 4), S/N Prefix <6021”
Getting Ready to Test
Before checking the assemblies, you must open the analyzer.
1.
Turn off the analyzer power.
2.
Unplug the power to the analyzer and disconnect all front and rear panel
connections except installed jumpers.
Equipment Type
Model or
Part Number
Alternate Model or Part
Number
Spectrum analyzer
8565E
856xE
a
a. Must be capable of measuring a signal at 7.61 MHz (analyzers with DSP version 4.0), or 7.44 MHz
(analyzers with DSP version 5.0), and 1 GHz.
Use an antistatic work surface and wrist strap to reduce the chance of
electrostatic discharge for all of the procedures in this chapter.
Содержание N5224B
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...12 Keysight N5224B 25B Service Guide Contents...
Страница 84: ...3 52 Keysight N5224B 25B Service Guide Tests and Adjustments Adjustments 3...
Страница 145: ...Keysight N5224B 25B Service Guide 4 61 Troubleshooting This page intentionally left blank...
Страница 212: ...5 50 Keysight N5224B 25B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 17 4 Port Receiver Group...
Страница 213: ...Keysight N5224B 25B Service Guide 5 51 Theory of Operation Receiver Group Operation Figure 5 18 2 Port Receiver Group...
Страница 220: ...5 58 Keysight N5224B 25B Service Guide Theory of Operation Power Supply Group Operation 5...
Страница 225: ...Keysight N5224B 25B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure...
Страница 490: ...6 270 Keysight N5224B 25B Service Guide Replaceable Parts Replaceable Parts Listings 6...
Страница 608: ...8 102 Keysight N5224B 25B Service Guide General Purpose Maintenance Procedures Correction Constants 8...
Страница 612: ...Appendix 4 Service Guide EEPROM Address Assignments and Location N5224 5A B PNA and N5244 5A B PNA X Instruments...
Страница 613: ...Installation Note Xxxxx xxxxx 3...