![Keysight N4431D Скачать руководство пользователя страница 102](http://html1.mh-extra.com/html/keysight/n4431d/n4431d_reference-manual_1961141102.webp)
5-12 Electronic Calibration Modules Reference Guide
Specifications and Characteristics
Characteristic Performance
5-
N4431D (7-16)
The characteristic performance in
the following table applies to N4431D (
if any of
the ports consist of 7-16 connector
). The data describes performance when
measuring “thru p
ath” A-B, C-D, A–D, A-D, B–C, and B-D.
N4431D (Type-N 50
Ω
)
The characteristic performance in the following table applies to N4431D
Option 020 (Type-N 50
Ω
female connectors on all ports). The data describes
performance when measuring “thru path” A-B, C-D, A–D, A-D, B–C, and B-D.
Table 5-21 Characteristic Performance for N4431D (m/f 7–16 connectors)
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz
5 GHz to 7.5 GHz
Transmission
Tracking (±dB)
b
0.18
0.05
0.08
Load Match (dB)
b
35
42
40
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance indicated in this
table.
b. Values based on using the network analyzer N5232B 400.
Table 5-22 Characteristic Performance for N4431D (Type-N 50
Ω
) Option 020
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz
5 GHz to 9 GHz
9 GHz to 13.5 GHz
Transmission
Tracking (±dB)
b
0.18
0.05
0.08
0.12
Load Match (dB)
b
35
43
41
38
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance indicated in this
table.