Making Measurements
2
To Measure Charge on an Infiniium Oscilloscope
N2820/1A User’s Guide
33
To make a Charge
measurement
1
Connect the probe. If an N2825A user-defined head is used, enter
the
R
SENSE
resistance as explained in
“Step 5. Configuring the Probe
2
If desired, click the oscilloscope’s
Zoom
button to turn on
zoom mode.
3
Click
Measure
>
Mixed
>
Charge (N282XA)
to view the Enter
Measurement Info dialog box.
Figure 13
Enter Measurement Info Dialog Box
4
If Zoom Mode is turned on, select
Gate to Zoom Window
to enable a
reduced zoom-window time span. If Zoom Mode is turned off, this
field is not selectable (grayed out).
5
If more than one N2820/1A probe is connected in the oscilloscope,
the dialog box allows you to select the probe used for the
measurement.
NOTE
With Zoom Mode turned on, you can use gating to show a reduced
zoom-window time span compared to the normal window. This is
indicated on the non-zoomed waveform by a white boundary-area box.
NOTE
Zoom mode is a software only expansion and results in the same vertical
resolution and accuracy between the zoom and non-zoom waveforms.
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...
Страница 110: ...This information is subject to change without notice Keysight Technologies 2018 March 2018 N2820 97006 www keysight com...