Safety Precautions

Safety Precautions

WARNING

This high voltage probe is designed to prevent accidental shock to 

the operator when properly used. This user guide must be read and 

understood prior to using the probe. Improper procedures or 

incorrect analysis of the measurement situation can result in serious 

shock.

If the probe is subjected to voltages over 15kV (DC +

 

AC peak) or 

10 kV rms then DO NOT USE. The probe could potentially be 

damaged. Return the probe immediately to a Keysight service 

center for performance validation.

WARNING

This high voltage probe is designed for use within Measurement 

Category I (CAT I) only. Do not use the probe for measurements 

performed on circuits defined by any other measurement category or 

transient overvoltages of more than 1500V. Refer to 

“IEC 

Measurement Category Definitions"

on page 6 for a definition of 

measurement categories.

WARNING

Do not exceed 60 seconds of contact between the probe and 

circuit-under-test when measuring voltages 

 8 kV (DC) or 

  5 kV 

(AC, rms). Keysight recommends that an interval of 5 minutes be 

taken between measurements.

WARNING

This high voltage probe must only be used by personnel who are 

trained, experienced, or otherwise qualified to recognize hazardous 

situations and who are trained in the safety precautions that are 

necessary to avoid possible injury when using such a device.

WARNING

Do not work alone when working with high voltage circuits.

Содержание N2771B

Страница 1: ...Keysight N2771B 15 kV High Voltage Probe User s Guide...

Страница 2: ...odify distribute or disclose the Software The EULA and the license set forth therein does not require or permit among other things that Keysight 1 Furnish technical information related to commercial c...

Страница 3: ...and nominal input capacitance between 6 and 20 pF The N2771B is a 1000 1 divider which extends the voltage measurement capability to maximum 15 kV DC AC peak 10 kVrms AC or 30 kVpeak 20 ns pulse widt...

Страница 4: ...ory I CAT I only Do not use the probe for measurements performed on circuits defined by any other measurement category or transient overvoltages of more than 1500V Refer to IEC Measurement Category De...

Страница 5: ...nnecting the probe WARNING The probe body should be kept clean and free of any conductive contamination Refer to the section on cleaning WARNING The ground lead is critical to the safe operation of th...

Страница 6: ...II is for measurements performed on circuits directly connected to the low voltage installation Example Household appliances portable tools and similar equipment Measurement Category III CAT III Measu...

Страница 7: ...r than the oscilloscope waveform if the probe attenuation on the oscilloscope has been set to 1 1 8 Turn off the high voltage source 9 Disconnect the Keysight N2771B High Voltage Probe from the high v...

Страница 8: ...8 kV DC or 5 kV AC rms Keysight recommends that an interval of 5 minutes be taken between measurements WARNING This high voltage probe is designed for use within Measurement Category I CAT I only Do...

Страница 9: ...imately 10 volts amplitude 4 Adjust the oscilloscope time base for 20 s div 5 Use a trimmer tool to adjust the compensation capacitor for the flattest pulse top See the figure below for the location o...

Страница 10: ...tio For input capacitance of 6 pF to 20 pF 1000 1 Bandwidth Designed For Use In 50 MHz 3 dB POLLUTION DEGREE 2 Input Resistance Operating Temperature 100 M when terminated by 1 M 0 C to 50 C Cable Len...

Страница 11: ...Voltage Derating 11 Voltage Derating The following graph shows the voltage versus frequency derating curve for the Keysight N2771B High Voltage Probe Figure 2 Voltage Versus Frequency Derating Curve...

Страница 12: ...12 Contacting Keysight Contacting Keysight Information on contacting Keysight Technologies can be found at www keysight com find contactus...

Отзывы: