100
Keysight M9502A/M9505A AXIe Chassis User Guide
For Module Developers
Chassis Backplane Connections
Chassis Backplane Connections
M9502A Backplane
The photo below reveals the M9502A backplane, with modules removed from all
slots. The backplane provides Zone 2 connectors P20, P21, P23, P24 and Zone 1
connector J10. Connector designations are shown for instrument slot 1.
The backplane differs from the M9505A in that the 2-slot backplane does not
utilize P22
A typical module layout is shown below the backplane photo, with the mating
connectors J20 through J24 and P10.
Depending on module type, you may implement all or none of J20-J24 for use in
the 2-slot chassis. You will always need connector P10 to power the module.
Optionally, you may need connector:
-
J20 to use the timing and trigger buses
-
J23 to use the base fabric (channel 1) and data fabric (channels 1 and/or 2)
interfaces
-
J23 and J20 to use up to 22 local bus interface pairs
-
J24 to expand local bus use to 42 pairs
-
J24 and J21 to expand local bus use to 62 pairs
-
J22 if the module is to be used as an instrument hub for the M9505A
Logical Slot 2
ESM Slot
Logical Slot 1
Instrument Slot 2
Logical Slot 3
Zone 2
Zone 1
AXIe Module Rear (mating) Edge
Instrument Slot 1 (Hub Slot)
Содержание M9502A
Страница 1: ...Keysight M9502A 2 Slot and M9505A 5 Slot AXIe Chassis User Guide...
Страница 2: ......
Страница 30: ...20 Keysight M9502A M9505A AXIe Chassis User Guide Introduction Product Warranty...
Страница 68: ...58 Keysight M9502A M9505A AXIe Chassis User Guide Navigating the Web Interface Chassis Health Page...
Страница 74: ...64 Keysight M9502A M9505A AXIe Chassis User Guide Using the Soft Front Panel SFP Screens...
Страница 82: ...72 Keysight M9502A M9505A AXIe Chassis User Guide Features and Functions Shelf Management Functions...
Страница 118: ...108 Keysight M9502A M9505A AXIe Chassis User Guide For Module Developers Chassis Backplane Connections...
Страница 138: ...128 Keysight M9502A M9505A AXIe Chassis User Guide Glossary...
Страница 141: ......