M9484C Startup Guide
21
Quick Start
Generate and View an Output Signal
Generate and View an Output Signal
After verifying the operation of the M9484C VXG, you are ready to make a
measurement.
1.
Turn on the Keysight signal analyzer.
2.
Connect a cable between the M9484C front panel RF Out connector and
the RF In connector on the Keysight signal analyzer.
3.
On the M9484C VXG, set the following parameters:
a. Frequency
: 2 GHz
b. Power
: -10 dBm
c.
Select the channel
1
button to turn on channel 1.
d.
For two and four channel instruments only: In the top right corner of
the display, set
RF Out (All)
to
On
by selecting the switch.
4.
On the Keysight signal analyzer, select
Frequency
in the menu panel, then
AUTO TUNE
. Verify that you have a 2 GHz, -10 dBm signal as shown
below.
The following measurement uses a M9484C VXG to generate the 2 GHz
signal and Keysight Signal Analyzer to analyze it. You may use any
frequency depending upon the signal analyzer used.
For an accurate frequency measurement, connect a 10 MHz Ref cable
between the M9484C and the signal analyzer.
For an accurate frequency measurement, connect a 10 MHz Ref cable
between the M9484C and the signal analyzer
.