Performance Verification Tests
Measurement Uncertainty
The insertion loss measurement is measured directly by the Agilent N5245A VNA. The VNA’s accuracy, when
measuring insertion loss, is specified to be ± 0.2 dB. For the purposes of this test, this is the expanded meas-
urement uncertainty.
If you use a VNA other than the Agilent N5245A (such as the other potential VNAs listed in the equipment list),
you will need to download the uncertainty calculator to calculate (worse-case) the uncertainty values for the meas-
urements you record in the
(see page
):
www.agilent.com/find/na_calculator
Insertion Loss below 10 MHz
Test Method
This test measures the insertion loss (below 10 MHz) of the M9360A, from the RF IN port to the RF 1 OUT port
and the RF 2 OUT port when in the preselector bypass setting. This test also measures the insertion loss (below 10
MHz) from the LO IN port to the LO 1 OUT and LO 2 OUT ports.
Limitations and Considerations
Methodology
This measurement is made using a signal source, two-resistor splitter and power meters/power sensors.
Specification
Table 1
Insertion Loss
Conditions
Source
≤ 2 dB
RF IN to RF 1 OUT, 100 kHz to 2.9 GHz, Minimum Attenuation
≤ 7 dB
RF IN to RF 2 OUT (Bypass), Minimum Attenuation
≤ 1 dB
LO IN to LO 1 OUT,≤ 10 GHz
≤ 1 dB
LO IN to LO 2 OUT,≤ 10 GHz
Equipment
Table 2
Equipment
Supported Models
Notes
Signal Source
Agilent E8267D; E8257D,
Power Meter
Agilent N1914A; N1912A; E4419B
Two-resistor splitter
Agilent 11667A (Opt 001/002)
Power Sensor (2 each)
Agilent N8482A; E9304A; 8482A
Adaptor: 3.5 mm (m) to Type-N (m)
1250-1743
Splitter to 1250-1748
Adaptor: 3.5 mm (m) to Type-N (f)
1250-1750
Sensor to M9360A
Adaptor, right angle SMA
1250-1397
Sensor to M9360A
Adaptor, 3.5 mm (m) to 3.5 mm (m) (qty2)
1250-1748; 83059A; 1250-1159
Sensor to M9360A
Agilent M9360A PXI Attenuator/Preselector Service Guide
29