14
Module Installation
Install the Module
Install the Module
Before installing the CXA-m module into a PXI express chassis, please read all the
safety information carefully and make sure the chassis is in a proper condition as
below.
• Ensure proper chassis air flow is maintained. The M9018A chassis has multiple
air intakes located on the lower sides, lower front and bottom front of chassis.
• Select a chassis that provides thermal protection if fans become inoperable or
forced air cooling is obstructed
• Use slot blockers and EMC filler panels in empty module slots to ensure proper
operating temperatures. Keysight chassis and slot blockers optimize module
temperature performance and reliability of test.
• Set chassis fans to high or auto. Do not disable fans.
• Position chassis to allow plenty of space around chassis air intake and fan
exhaust.
• At environment temperatures above 45°C, set chassis fan speed to high.
WARNING
Keysight Technologies does not warrant third-party system-level (combination of
chassis, controllers, modules, etc.) performance, safety, or regulatory compliance,
unless specifically stated.
CAUTION
PXI hardware does not support "hot-swap" (changing modules while power is applied to
the chassis) capabilities. Before installing or removing a module to/from the chassis,
power off the chassis to prevent damage to the module.
The CXA-m module in a PXI express chassis occupies 4 empty slots. Please refer to
the chassis documentation for further information about how to use the chassis.
Perform the following steps to install a CXA-m module in a chassis:
1. Make sure the line cord is plugged in to establish earth ground and the chassis
power switch is in the Off (Standby) position.
2. If the chassis has multiple fan speed settings, ensure the fans are set to automatic.
Do not set the fan speed to low or turn it off.
Содержание M9290A
Страница 18: ...18 Module Installation Instrument Symbols...
Страница 34: ...34 Software Installation and Licensing Transporting a License...
Страница 68: ...68 Running Modular SA application TCP IP Address Configuration...
Страница 74: ...74 Troubleshooting Returning an Analyzer for Service...
Страница 75: ......
Страница 76: ...Keysight Technologies Inc 2014 Printed in China October 2014 M9290 90007 M9290 90007 www keysight com...