Test 7: AC Current Verification Test
Keysight M9182A/M9183A Calibration and Verification Procedure
31
Test 7: AC Current Verification Test
1
Make sure you have read
2
Connect the calibrator to the input terminals of the DMM, as shown below.
3
Set the AC Current function on the M9182A/83A DMM.
4
Select each range in the order shown below. Using the calibrator, provide the
input current and frequency indicated. Compare AC current measurement
results from the M9182A/83A DMM to the appropriate test limits shown in the
table. Be certain to allow for appropriate source settling.
Calibrator
input ACI (A)
Calib.
freq. (Hz)
DMM
function
DMM
range (A)
Quick
check
Error from nominal (± A)
24 hours
90 days
1 year
60.0000E-06
10.0000E+00
ACI
2.000000E-03
17.620E-06
17.740E-06
17.740E-06
2.30000E-03
20.0000E+00
ACI
2.000000E-03
24.700E-06
24.700E-06
27.000E-06
60.0000E-06
47.0000E+00
ACI
2.000000E-03
Q
13.620E-06
15.050E-06
16.070E-06
2.30000E-03
1.0000E+03
ACI
2.000000E-03
2.520E-06
4.840E-06
6.760E-06
60.0000E-06
10.0000E+03
ACI
2.000000E-03
16.070E-06
16.08E-06
16.13E-06
20.0000E-03
47.0000E+00
ACI
20.00000E-03
24.00E-06
50.00E-06
62.00E-06
200.0000E-03
1.0000E+03
ACI
200.0000E-03
Q
460.0E-06
520.0E-06
620.0E-06
2.0000E+00
1.0000E+03
ACI
2.00
9.800E-03
10.400E-03
11.000E-03