Keysight PXIe System and Cable Interface Installation Guide
23
M9021A PCIe Cable Interface Module Installation
M9021A PCIe Cable Interface Module Installation
The M9021A uses special features built into the Keysight M9018 chassis and is
not compatible with the PXIe System Slot in other PXIe chassis.
M9021A Switch Settings
Switch
Description
S201
Ensure that S201 is in the NC position (right side down).
S301
For applications, where the M9021A provides a cabled PCIe link
between the M9018A chassis and an external host computer, install
the M9021A in the system slot (slot 1) of the M9018A PXIe chassis.
The two switches on S301 must be in the Target (right) position. This
is the default setting.
For connecting to a downstream RAID solution, such as the JMR
mass storage, the M9021A must be installed in a x8 hybrid slot
(either slot 2, 6, 11, or 15) in the PXIe chassis. In this case, both
switches on S301 should be in the Host (left) position.
Figure 6
M9021A Switch Locations
S301. Switches shown in Target
(default) position; used when the
M9022A is installed in system slot
(slot 1) of the M9018A chassis.
Move both switches to Host (left)
position for use with the M9392A
RAID array (but only in a data
streaming/storage solution).
S201. Ensure that S201 is in the
NC position, (right side down).