132
Keysight M8121A Arbitrary Waveform Generator User Guide
6
Remote Programming
:CaLibrate Commands
The
:CALibrate
commands are used to control the latency from the
M8131A digitizer to the M8121A AWG. This set of commands should be
used only when controlling both instruments. The end effect is that the
delay from the analog input of the digitizer to the analog output of the
AWG is a fixed latency for a given setup. The calibration procedure must
be run upon a change in conditions, such as a disruption to the data flow
over ODI, change in cables, or a change in clock source.
:CALibrate:SYNC:INPut:LEVel
Command
:CAL:SYNC:INP:LEV <level>
Query
:CAL:SYNC:INP:LEV?
Parameters
<level>|MINimum|MAXimum|DEFault
Parameter Suffix
None
Description
Set or query the Sync input threshold in Volts.
The Sync Input is used with the M8131A digitizer to calibrate a
deterministic latency from analog input to analog output.
Examples
:CALibrate:SYNC:INPut:LEVel 0.4
:CALibrate:LATency:LCMP
Command
:CALibrate:LATency:LCMP <lcmp_value>
Parameters
<lcmp_value>
Parameter Suffix
None
Description
Lowest common multiplier between the internal clocks of the M8131A and
M8121A. Use value of 5 when operating with a M8131A digitizer.
Examples
:CALibrate:LATency:LCMP 5
Содержание M8121A
Страница 1: ...User Guide Keysight M8121A Arbitrary Waveform Generator ...
Страница 20: ......
Страница 42: ...42 Keysight M8121A Arbitrary Waveform Generator User Guide 3 Soft Front Panel Figure 7 M8121A startup screen ...
Страница 72: ......
Страница 138: ......
Страница 142: ......