Characteristics and Specifications
1
2000/3000 X-Series Oscilloscopes Service Guide
15
Transient Withstand Capability
CAUTION
Maximum input voltage for analog inputs
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
with N2862A/B or N2863A/B 10:1 probe: 300 Vrms
CAUTION
Do not exceed 5 Vrms in 50
Ω
mode (3000 X-Series oscilloscopes only). Input protection
is enabled in 50
Ω
mode, and the 50
Ω
load will disconnect if greater than 5 Vrms is
detected. However, the input could still be damaged, depending on the time constant of
the signal. The 50
Ω
input protection mode only functions when the oscilloscope is
powered on.
CAUTION
Maximum input voltage for logic channels:
±40 V peak CAT I; transient overvoltage 800 Vpk